TECHNICAL REFERENCE___________________________________________________________

34 __________________________________________________________________ M211097EN-C

Table 10

Required Parameters for WS425 Analog Output
Operation Mode

Parameter Name Default

Value

Allowed Values

Description

aoErrValue

1000

0 …. 32000

Set 2 for voltage output and 1000 for
frequency output for similar operation as
with WS425.

aout1_g

1

0 …. 100

Gain for AOUT1.
Set 0.01789 for compatibility with WS425
voltage output.
Set 11.18 for compatibility with WS425
frequency output.

aout2_g

1

0 …. 100

Gain for AOUT2.
Set 0.0027 for compatibility with WS425
potentiometer output.

aout1mode 3

(0

=

current)

1 = voltage
2 = frequency
(3 = disabled)

Analog output mode for AOUT1.
Set 1 for voltage and 2 for frequency.

aout2mode 7

(4

=

current)

(5 = voltage)
6 = potentiometer
(7 = disabled)

Analog output mode for AOUT2.
Set 6 for potentiometer.

aout1_o
aout2_o

0

-10000 … 10000

Offset for AOUT1 and AOUT2.
Set output offset to 0.

For configuration instructions, see Vaisala WINDCAP

®

Ultrasonic Wind

Sensor Series WMT700 User's Guide.