WS425 F/G ASOS Data Message

WMT700 responds to the WA command with a data message described
in Table 13 below.

Table 13

WS425 F/G ASOS Data Message

Byte Description
1

Start of text

2 Sensor

ID

3 Command

identifier

4 Sensor

status:

P = Pass
F = Fail
H = Heater failure
For further information, refer to notes 1 and 2.

5-7

Averaged wind direction (degrees)

8-10

Wind direction (degrees) @ peak wind speed

11-12

Averaging time for wind speed and direction

13-14

Averaging time for peak wind speed and direction (seconds)

15-19

Averaged wind speed

20-24

Peak wind speed since last WA command

25

Wind speed unit:
M = miles per hour
K = knots
L = kilometers per hour
T = meters per second

26-27

Signal quality. For further information, refer to note 3.

28-29

Hex checksum. For further information, refer to note 4.

30

End of text

31 Carriage

return

32 Line

feed

Notes:

1.

The sensor status (byte 4) indicates the following:
- P (Pass) indicates that all diagnostic tests have passed and

WMT700 is functioning normally.

- F (Fail) indicates that one or more of the diagnostic tests have

failed. You can then request diagnostics and extended tests
(WS) to further isolate the problem.

- H (Heater failure) indicates that the wind reading is valid but

one or more built-in heater diagnostic tests have failed.

2.

When the status byte is F, the wind direction fields are set to 999
and the wind speed fields are set to 999.9 in the applicable sensor
response messages.

3.

The signal quality (bytes 26-27) is a numeric value from 0 to 99
representing the data quality of processed data samples during the
averaging time.