Chapter 5 ______________________________ Operating WMT700 with WS425 and SDI-12 profiles

VAISALA _______________________________________________________________________ 65

aAb! — Change Address

This command changes the WMT700 address. After WMT700 has
received this command and sent a response, the sensor does not respond
to any other command for one second. This gives WMT700 time to write
the new address to the non-volatile memory.

NOTE

You can also change the WMT700 address by entering configuration
mode and changing the address parameter.

aAb!

where
a =

The

current

single-digit WMT700 address that

corresponds to the first character of the address value.

A

= The change address command.

b

= The new address.

!

= Terminates the command.

The response is the following:

b<CR><LF>

where
b

= The new single-digit WMT700 address (or the original

address if WMT700 is unable to change the address).

<CR><LF> = Terminates the response.

aC! — Start Concurrent Measurement

This command starts concurrent measurement. WMT700 does not return
measurement data after this command. It sends a response that indicates
when the measurement result can be retrieved using the aD0! command
and the number of parameters.

aC!

where
a

= The single-digit WMT700 address that corresponds to

the first character of the address value.

C

= The start concurrent measurement command.

!

= Terminates the command.