Aalborg PWE Digital User Manual

Page 45

Advertising
background image

40

4.6.2 Main Totalizer Flow Start Numerical entry

The Main Totalizer Start Flow value can be entered in increments of 0.1% from 0.0
– 100.0%F.S. A typical display with flow meter Main totalizer Start Flow settings is
shown below.

4.6.3 Main Totalizer Event Volume Numerical entry

Main Totalizer Event Volume value must be entered in currently active volume /
mass based engineering units. Totalizer action event become true when the total-
izer reading and preset "Event Volume" values are equal.

If the Totalizer Event at preset total volume feature is not required, then set
Totalizer Event Volume value to zero (default settings).

A typical display with flow meter Main Totalizer Event Volume settings is shown
below.

4.6.4 Main Totalizer Reset Tabular entry

The Main Totalizers reading can be reset by pressing ENTER button. A typical dis-
play with flow meter Main Totalizer Reset screen is shown below.

The next conformation screen will appear only for 2 seconds.

MT Flow Start

5.0 %F.S.

MT Event Volume

0.0 Gal

Reset Main Total

Value?

Press Ent key to

reset Totalizer!