54

Figure 5-1 PWE configuration jumpers

Table 5-1

Function

Analog Flow Output Analog Temp. Output* RS-485

J3A

J3B

J3C

J3D

J3E

J3F

J3G

Analog

Output

0-5 Vdc

2-3

5-6

8-9

11-12 14-15 17-18

4-20 mA

1-2

4-5

7-8

10-11 13-14 16-17

RS-485

220 Ohm

Termination

Resistor

OFF

20-21

ON

19-20

5.2

Analog Output Signals calibration

Note: The analog output available on the PWE Flow Meter was calibrated
at the factory for the specified fluid and full scale flow range (see the
device’s front label). There is no need to perform analog output
calibration unless the EEPROM IC was replaced or offset/span
adjustment is needed. Any alteration of the analog output scaling
variables in the EEPROM table will VOID calibration warranty supplied
with instrument.

,