Aalborg PWE Digital User Manual

Page 70

Advertising
background image

65

COMMAND

NAME

DESCRIPTION

NO.

COMMAND SYNT

A

X

COMMAND

ARGUMENT 1

ARGUMENT 2

ARGUMENT 3

ARGUMENT 4

RESPONSE

Te

mperature

Alarms**

Sets / reads the status of the

temperature alarms.

Note: High and Low limits

have to be entered in the

°C.

High alarm value has to be

more than Low alarm value.

Alarm conditions:

Temp. > High Limit = H

Temp. < Low Limit = L

Low < T

emp. < High = N

7

TA

H (high flow limit)

<V

alue>

(-10.1-100

°C)

TA

H:<V

alue (

°C)>

L (low flow limit)

<V

alue>

(-10.1-100

°C)

TA

L:<V

alue (

°C)>

A (action delay in

seconds)

<V

alue> (0-3600

sec.)

TAA:<V

alue (sec)>

E (enable alarm)

TA

:E

D (disable alarm)*

TA

:D

R (read current

status)

TA

:N (no alarm)

TA

:H (high alarm)

TA

: L (low alarm)

S (Read current

settings)

TAS:M,L,H,D,B where:

M – mode (E/D)

L – Low settings (

°C)

H – High settings (

°C)

D – Action Delay (sec)

B – Latch mode (0-3)

B

B

lo

ck

(

L

at

ch

)

mode

<V

alue>

(0-disabled*)

(1-enabl’d L)

(2-enabl’d H)

(3 –both L,H)

TAB:<V

alue> where:

V

alue = 0 – 3