Aalborg PWE Digital User Manual
Page 70
65
COMMAND
NAME
DESCRIPTION
NO.
COMMAND SYNT
A
X
COMMAND
ARGUMENT 1
ARGUMENT 2
ARGUMENT 3
ARGUMENT 4
RESPONSE
Te
mperature
Alarms**
Sets / reads the status of the
temperature alarms.
Note: High and Low limits
have to be entered in the
°C.
High alarm value has to be
more than Low alarm value.
Alarm conditions:
Temp. > High Limit = H
Temp. < Low Limit = L
Low < T
emp. < High = N
7
TA
H (high flow limit)
<V
alue>
(-10.1-100
°C)
TA
H:<V
alue (
°C)>
L (low flow limit)
<V
alue>
(-10.1-100
°C)
TA
L:<V
alue (
°C)>
A (action delay in
seconds)
<V
alue> (0-3600
sec.)
TAA:<V
alue (sec)>
E (enable alarm)
TA
:E
D (disable alarm)*
TA
:D
R (read current
status)
TA
:N (no alarm)
TA
:H (high alarm)
TA
: L (low alarm)
S (Read current
settings)
TAS:M,L,H,D,B where:
M – mode (E/D)
L – Low settings (
°C)
H – High settings (
°C)
D – Action Delay (sec)
B – Latch mode (0-3)
B
B
lo
ck
(
L
at
ch
)
mode
<V
alue>
(0-disabled*)
(1-enabl’d L)
(2-enabl’d H)
(3 –both L,H)
TAB:<V
alue> where:
V
alue = 0 – 3