5.2.4 Temperature 0-5 Vdc analog output

calibration*

1. Install jumpers J3D, J3E and J3F on the PC board for 0-5 Vdc output

(see Table 5-1).

2. Connect a certified high sensitivity multi meter set for the voltage

measurement to the pins L (+) and K (-) of the PWE 12 Pin "M16"
connector.

3. Write 4000 counts to the DAC_B channel: !11,WRITE,1,4000[CR]
4. Read voltage with the meter and calculate TOutScaleV value:

5. Save TOutScaleV in to the EEPROM: !11,MW,43,X[CR]

Where: X – the calculated ToutScaleV value.

5.2.5 Temperature 4-20 mA analog output

calibration*

1. Install jumpers J3D, J3E and J3F on the PC board for 4-20 mA output

output (see Table 5-1).

2. Connect a certified high sensitivity multi meter set for the current

measurement to pins L (+) and K (-) of the PWE 12 Pin "M16" connector.

3. Write 4000 counts to the DAC_B channel: !11,WRITE,1,4000[CR]
4. Read current with the meter and calculate ToutScale_mA value:

5. Write zero counts to the DAC_B channel: !11,WRITE,1,0CR]
6. Read offset current with the meter and calculate ToutOffset_mA value:

7. Save ToutScale_mA in to the EEPROM: !11,MW,45,Y[CR]

Save ToutOffset_mA in to the EEPROM: !11,MW,46,Z[CR]

ToutScaleV=

20000

Reading[V]

ToutScale_mA=

4000

Reading[mA]

ToutOffset_mA=-TOutScale_mA*Offset_Reading[mA]

Advertising