Myron L PS6FCE User Manual

Page 11

Advertising
background image

6

7

III.

RULES of OPERATION

A. Operation

Using the instrument is simple:

Individual or multiple parameter readings may be obtained by

filling individual sensors or entire cell cup area.

Rinse the conductivity cell or pH/ORP sensor well with test

solution 3 times and refill. Temperature and/or measurement

extremes will require additional rinses for maximum accuracy.

Press the desired measurement key to start measurement.

Pressing the key again does no harm and restarts the 15

second auto “off” timer.

Note the value displayed or press the MS key to store the

reading (ref. Memory Storage, pg. 23). It’s that simple!

B. Characteristics of the Keys

Though your P

ool

P

ro

has a variety of sophisticated options,

it is designed to provide quick, easy, accurate measurements

by simply pressing one key.

All functions are performed one key at a time.

There is no “off” key. After 15 seconds of inactivity the

instrument turns itself off (60 seconds in CAL mode). User

adjustable up to 75 seconds.

Rarely is it necessary to press and

hold a key (as in Procedure

to Select a Solution, pg. 15; or Cond. MIN/SALT or TDS

Calibration, pg.17).

C. Operation of the Keys (See Instrument Illustration on pg. i)

1. Measurement Keys in General

Any of the measurement keys in the upper part of the keypad turns on

the instrument in the mode selected. The mode is shown at the bottom

of the display, and the measurement units appear at the right. Pressing

a measurement key does this even if you are in a calibration sequence

and also serves to cancel a change (ref. Leaving Calibration, pg. 17).

2. COND, MIN/SALT and TDS Keys

These 3 keys are used with solution in the Conductivity Cell.

Precautions:

While filling cell cup ensure no air bubbles cling on the cell wall.

If the proper solution is not selected (KCl, NaCl, 442),

refer to Why Solution Selection is Available, pg. 14 and

Procedure to Select a Solution, pg. 15.

a. COND Key

Solution to be tested is introduced into the conductivity cell and a press

of

displays conductivity with units on the right. On the left is

shown the solution type selected for conductivity.