8

9

5. UP or DOWN Keys

While measuring in any parameter, the

or

keys activate

the Memory Store and Memory Recall functions.

While in CAL mode, the keys step or scroll the displayed value up or

down. A single press steps the display and holding either key scrolls the

value rapidly.

While in Memory Recall, the keys scroll the display up and down through

the stack of records (ref. Memory Recall, pg. 23).

IV.

AFTER USING the P

ool

P

ro

A. Maintenance of the Conductivity Cell

Rinse out the cell cup with clean water. Do not scrub the cell. For oily

films, squirt in a foaming non-abrasive cleaner and rinse. Even if a

very active chemical discolors the electrodes, this does not affect the

accuracy; leave it alone. (ref. Cleaning Sensors, pg. 32)

B. Maintenance of the pH/ORP Sensor

The sensor well must be kept wet with a solution. Before replacing

the rubber cap, rinse and fill the sensor well with Myron L pH Sensor

Storage Solution. If unavailable, you can use an almost saturated KCl

solution, pH 4 buffer or at least a strong table salt solution. NEVER

USE DISTILLED WATER. (ref. pH and ORP Practices, pg. 22).

V.

SPECIFIC RECOMMENDED MEASURING

PROCEDURES

If the proper solution is not selected (KCl, NaCl, 442), see Solution

Selection, pg. 14.

NOTE: After sampling high concentration solutions or temperature

extremes, more rinsing may be required. When sampling low conductivity

solutions, be sure the pH cap is well seated so that no solution washes

into the conductivity cell from around the pH cap.

A. Measuring Conductivity MIN/SALT & Total Dissolved Solids

(TDS)

1.

Rinse cell cup 3 times with sample to be measured. (This

conditions the temperature compensation network and

prepares the cell.)

2.

Refill cell cup with sample.

3.

Press

,

MIN

SALT

or

.

4.

Take reading. A display of [- - - -] indicates an overrange

condition.