XII. TOTAL RETURN to FACTORy SETTINGS “FAC SEL”

There may come a time when it would be desirable to quickly reset

all the recorded calibration values in the instrument back to the factory

settings. This might be to ensure all calibrations are set to a known

value, or to give the instrument to someone else free of adjustments or

recorded data for a particular application.

NOTE: All stored data will be lost.

1.

Press

.

2.

Press

to display the stored memory records.

3.

Press

repeatedly until

you pass the CL

r

ALL

and the C-F locations. The

display will show a

FAC SEL” (see Figure 19).

4.

Press

to accept the resetting. Display will return to

Conductivity mode.

XIII.

CELL CHECK

The cell check verifies the cleanliness of the conductivity/TDS/MIN

/

SALT

sensor. In normal use the cell may become dirty or coated and require

cleaning. If the display is showing “.00” when the cell cup is dry, the sensor

is probably clean. No matter what a manufacturer claims, a sensor can

and will become contaminated or coated; therefore require cleaning. A

true 4-wire sensor, as in the P

ool

P

ro

, helps to mitigate contamination,

however, NO SENSOR IS 100% IMMUNE.