D. A Chart of Comparative Error

In the range of 1000 µS, the error using KCl on a solution that should be

compensated as NaCl or as 442, is illustrated in the Chart 2 below.

Users wanting to measure natural water based solutions to 1% would

have to alter the internal compensation to the more suitable preloaded

“442” values or stay close to 25°C. Users who have standardized to

KCl-based compensation may want to stick with it, regardless of

increasing error as you get further from 25°C. The P

ool

P

ro

will provide

the repeatability and convertibility of data necessary for relative values

for process control.

E. Other Solutions

A salt solution like sea water acts like NaCl. An internal correction for NaCl

can be selected for greatest accuracy with such solutions. Many solutions

are not at all similar to KCl, NaCl or 442, however, are still referenced to

one of these for the purpose of commonality.

Clearly, the solution characteristics should be chosen to truly represent

the actual water under test for rated accuracy of ±1%. Many industrial

applications have always been relative measurements seeking a number

to indicate a certain setpoint or minimum concentration or trend. The

P

ool

P

ro

gives the user the capacity to take data in the “KCl conductivity

7%

Chart 2

55

(1)%

(2)%

0%

1%

2%

3%

4%

5%

6%

0 5 10 15 20 25 30 35 40 45 50

Temperature

NaCl error with KCl tempco

442 error with KCl tempco