The problem is, on the other side of the membrane is an unknown test

solution, not potassium chloride. The outside electrode, also called

the Reference Junction, is of the same construction with a porous

plug in place of a glass barrier to allow the junction fluid to contact the

test solution without significant

migration of liquids through the

plug material. Figure 30 shows

a typical 2 component pair.

Migration does occur, and this

limits the lifetime of a pH junction,

from depletion of solution inside

the reference junction or from

contamination. The junction may

be damaged if dried out because

insoluble crystals may form in a

layer, obstructing contact with test

solutions. See pH/ORP, pg. 42.

4. The Myron L Integral pH Sensor

The sensor in the P

ool

P

ro

(see

Figure 31) is a single construction

in an easily replaceable package.

The sensor body holds an oversize

solution supply for long life.

The reference junction “wick” is

porous to provide a very stable,

low permeable interface, and is

located under the glass pH sensing

electrode. This construction

combines all the best features of

any pH sensor known.

5. Sources of Error

The basics are presented in

pH/ORP, pg. 42.

a. Reference Junction

The most common sensor problem will be a clogged junction because

a sensor was allowed to dry out. The symptom is a drift in the “zero”

setting at 7 pH. This is why the P

ool

P

ro

does not allow more than 1 pH

unit of offset during calibration. At that point the junction is unreliable.

b. Sensitivity Problems

Sensitivity is the receptiveness of the glass surface, which can be

diminished by a film on the surface. This problem also causes long

response time.

Glass surface

Figure 30

KCl solution

Electrode wire

Electrode

wire

H

+

ions

Junction

Plug

KCl solution

Figure 31

Junction plug

Platinum button

H

+

ions

Electrode wires

Glass

Glass

Surface