Myron L PS6FCE User Manual

Page 18

Advertising
background image

14

15

4. FC

E

Best Practices

For best results it is recommended that you:

1.

Take 3 consecutive FC

E

measurements and record the

readings.

2.

Calculate the average of the 3 measurements. Use this value.

3.

Ignore measurements that are significantly different from the

others. Ex: 3.20 ppm, 1.15 ppm, 3.10 ppm

IMPORTANT: After use, fill pH/ORP sensor well with Myron L pH

Sensor Storage Solution and replace protective cap. If Myron L pH

Sensor Storage Solution is unavailable, you can use a strong KCl

solution, a pH 4 buffer, or a saturated solution of table salt and tap

water (ref. Cleaning Sensors, 2. pH/ORP, pg. 32). Do not allow pH/

ORP sensor to dry out.

VI.

SOLUTION SELECTION

A. Why Solution Selection is Available

Conductivity, MIN/SALT, and TDS require temperature correction to

25°C values (ref. Standardized to 25°C, pg. 37). Selection determines

the temperature correction of conductivity and calculation of TDS from

compensated conductivity (ref. Cond. Conversion to TDS, pg. 40).

B. The 3 Solution Types

On the left side of the display is the salt solution characteristic used

to model temperature compensation of conductivity and its TDS

conversion. Generally, using KCl for Conductivity, NaCl for Mineral/

Salt, and 442 (Natural Water characteristic) for TDS will reflect present

industry practice for standardization. This is the setup as shipped from

the factory (ref. Solution Characteristics, pg. 40).

C. Calibration of Each Solution Type

There is a separate calibration for each of the 3 solution types. Note

that calibration of a 442 solution does not affect the calibration of a

NaCl solution. For example: Calibration (ref. Conductivity, MIN/SALT

or TDS Calibration, pg. 17) is performed separately for each type

of solution one wishes to measure (ref. Conductivity/TDS Standard

Solutions, pg. 36).

Advertising