Myron L 6Psi and 4P User Manual

Page 16

Advertising
background image

12

you wish to change the

solution type.

2.

Press and hold

key

for 3 seconds to make

SEL” appear (see Figure

1). For demonstration

purposes, all 4 solution

types are shown simultaneously.

3.

Use the

or

key to select type of solution desired

(ref. Solution Characteristics, pg. 40). The selected solution

type will be displayed: KCl, NaCl, 442 or User.

4.

Press

to accept new solution type.

E. Application of User Solution Type

1. User Programmable Temperature Compensation

(Tempco)

This feature allows you to change your Ultrameter II’s temperature

compensating factor to another factor between 0-9.99%/°C (ref.

Temperature Compensation, pg. 37). This feature does not apply to pH

or ORP.

a.

As in Procedure to Select a Solution, pg. 11, select “User”

mode.

b.

With “User” mode now selected, press

. You may now

adjust a temperature compensation from .00%/°C to 9.99%/°C,

by pressing

or

.

See example in Figure 2.

c.

Press

twice to skip

calibration adjustment and

accept the new tempco (3

times if in TDS mode). You

are now ready to measure

samples with your new temperature compensation factor.

Figure 1

KCl

442

NaCl

User

Figure 2

°C

% /

User