Myron L 6Psi and 4P User Manual

Page 52

Advertising
background image

48

XXVI. ADDENDUM: NEW ORP PPM FREE ChLORINE FEATURE

For 6Psi ONLy Software V.3.51 and later (see pg. 46).

Contact the Myron L Company for more information:

www.myronl.com

*Given water is sanitized by only chlorine at a pH of 5-9 and a temperature of 20-30ºC

A convenient new feature has been added to the Ultrameter II 6Psi

that allows the user to choose between ORP millivolts (mV) and parts per

million (ppm) free chlorine readings. mV and ppm free chlorine are the two

most commonly used sanitizer measurement units used in water quality

management. The 6Psi mV to ppm free chlorine conversion algorithm is

based on a published conversion curve that assumes a pH of 5-9 and a

temperature of 20-30ºC in water sanitized by chlorine only

and bench testing

performed at the Myron L Company. We have increased the accuracy of low-

range ppm free chlorine readings by extrapolating from the published data.

With this new feature the 6Psi can measure a dynamic range of sanitizer

concentrations that is wider than the range of a colorimetric test kit.

Plot Showing the Conversion Curve between ppm and mV*

Additional ORP Specifications

Spec

Value

Ranges

± 999 mV

0.20 – 9.99 ppm

Resolution

mV resolution: 1 mV

ppm resolution: 0.01 ppm

Accuracy

± 1 mV

± 2.5% of reading ppm*

8

10

6

4

2

0

0

100

200

600

500

400

300

700

800

ORP (mV)

ORP

(ppm

free

chlorine)

ORP mV to ppm Conversion Curve