Myron L 6Psi and 4P User Manual

Page 58

Advertising
background image

C.

hardness Unit Preference Selection

The LSI Calculator allows you to select either ppm or grains of

hardness units for saturation index calculation.

The hardness unit conversion is based on the following equivalency:

17.1 ppm (mg/L) = 1 grain

The hardness range is limited to 0.0 - 1710 ppm and 0.0 - 100 grains

of hardness in the 6Psi.

NOTE: The 6Psi MUST be in the

hardness screen (hrd) in LSI

Calculator mode or LSI Calculator

“User” mode for hardness unit

preference selection.

By default, hardness unit preference

is set to ppm (see Figure 51).

1.

While in the hrd screen in

CALC LSI or LSI Adj mode,

press and hold the

key. hArd SEL will display

along with the current unit

preference (see Figure 52).

2.

Press

or

to

toggle between ppm and

grains of hardness. No unit

icon is lit when grains is

selected (see Figure 53).

3.

Press

to accept the

change. The LSI calculator

will resume in the hrd

screen. The unit change

will be reflected on the

display (see Figure 54).

Figure 51

PPM

PPM

Figure 52

Figure 53

Figure 54