29

6.

Press

key to accept the setting. Pressing any of the

measurement keys will exit without changing the setting. “User”

mode “Linc” is now complete. The “User” mode will now use the

calibration gain constant used for the calibration of the

Standard Solution as outlined above.

C. Canceling “User” Mode Calibration “Linc”

The Ultrameter II must be in “User” linked mode in order to cancel the

“Linc” (ref. SOLUTION SELECTION, pg. 11).

1.

Press “Linked” measurement key

,

or .

Two solution icons will be shown in the left side of display -

User” and another, e.g., “KCl”.

2.

Press

key until the menu “Linc” appears (see Figure 27).

3.

Press

key; the instrument will display both “SEL” and

the “User” Icon.

4.

Press

key until “User” is the only solution icon being

displayed.

5.

Press

key.

6.

The “User” mode calibration “Linc” has now been canceled.

NOTES:

1. To maintain repeatability, use the same standard solutions for future

calibrations.

2. Calibration of the Ultrameter II Gain Factor for “User” mode is not

available when the calibration linc has been established. The other

calibration functions (i.e., Temperature Compensation %/C settings and

TDS Ratio settings) are still intact. To perform a calibration of the “User”

mode as described in User Calibration Conductivity/TDS, pg. 15, the

“User” mode Linc should be canceled. See above, “Canceling “User”

mode calibration “Linc””.

3. Once a “Linc” has been established for “User” mode , the “Linc” will

apply to all measurement modes using User solution selection (i.e.,

TDS/User, Cond/User or Res/User).