Myron L 6Psi and 4P User Manual

Page 31

Advertising
background image

27

2.

Press

to display the stored memory records.

3.

Press

repeatedly until you pass the CELL ch location.

The display will show “Auto oFF” (see Figure 24).

4.

Press

to initiate. “CAL

will be displayed along with

15 SEC” or current Auto

oFF value (see Figure 25).

5.

Press

or

to

change the amount of time

(see Figure 26). Maximum

time is shown.

6.

Press

to accept the

change (new time).

XV.

“USER” MODE CALIBRATION LINC™ FUNCTION

The Linc™ function allows easy calibration when in “User” mode and the

user does not have a user standard solution to calibrate the instrument.

This function will ensure more repeatable and accurate measurements

than many other calibration methods. It is recommended that this

function be used to provide the highest degree of confidence when the

Ultrameter II is used in “User” mode. When Linc is used, the “User”

mode is linked to another standard, i.e., if User and KCl are linked, a KCI

standard solution is used to calibrate the instrument. It is that simple.

A. Calibration of Ultrameter II for use in “User” mode

1.

Press

or

key.

2.

Calibrate the unit using a Standard Solution (ref.

CALIBRATION, pg. 14).

3.

Place the Ultrameter II in “User” mode (ref. SOLUTION

SELECTION, pg. 11).