Myron L 6Psi and 4P User Manual

Page 22

Advertising
background image

18

reduce your variation from factory setting by pressing

or .

6.

Press

to accept the new value. The pH Zero Calibration

is now complete. You may continue with pH gain Calibration or

exit by pressing any measurement key.

b. pH gain Calibration (6Psi)

Important: Always calibrate or verify your Ultrameter II with a pH 7

buffer solution before adjusting the gain with acid or base buffers, i.e., 4

and/or 10, etc. Either acid or base solution can be used for the 2nd point

“gain” calibration and then the opposite for the 3rd point. The display

will verify that a buffer is in the sensor well by displaying either “Acd” or

bAS”.

1.

The pH calibration mode is initiated by either completion of the

pH Zero Calibration, or verifying 7 buffer and pressing the

key twice while in pH measurement mode.

2.

At this point the “CAL”, “BUFFER” and “Acd” or “bAS

annunciators will be displayed (see Figures 7 and 8).

NOTE: If the “Acd” and “bAS” indicators are blinking, the unit is indicating

an error and needs either an acid or base solution present in the sensor

well.

3.

Rinse sensor well 3 times with acid or base buffer solution.

4.

Refill sensor well again with same buffer solution.

5.

Press

or

until display agrees with buffer value.

Figure 7

BUFFER

pH

CAL

Figure 8

pH

BUFFER

CAL

Advertising