Myron L 6Psi and 4P User Manual

Page 19

Advertising
background image

15

measurement key cancels changes not yet accepted and exits calibration

mode.

Leaving pH after the 2nd buffer results in the same gain being entered

in place of the 3rd buffer.

2. Calibration Limits

There are calibration limits. A nominal “FAC” value is an ideal value

stored by the factory. Attempts to calibrate too far, up or down, from there

will cause the displayed value to be replaced with “FAC”. If you accept it

(press the “Cal” key), you will have the original default factory calibration

for this measurement. The need to calibrate so far out that “FAC” appears

indicates a procedural problem, incorrect standard solution, a very dirty

cell cup or an aging pH/ORP sensor (ref. Troubleshooting Chart, pg.

34).

C. Calibration Procedures

1. Conductivity or TDS Calibration

a.

Rinse conductivity cell three times with proper standard (KCl,

NaCl, or 442) (ref. Cond/TDS Standard Solutions, pg. 36). For

user calibration see User Calibration Conductivity/TDS below.

b.

Refill conductivity cell with same standard. KCl-7000 shown.

c.

Press

or

, then

press ; “CAL” icon will

appear on the display

(see Figure 5).

d.

Press

or

to

step the displayed value toward the standard’s value (7032 >

7000) or hold a key down to scroll rapidly through the reading.

e.

Press

once to confirm new value and end the

calibration sequence for this particular solution type. If another

solution type is also to be measured, change solution type now

and repeat this procedure.

2. User Calibration Conductivity/TDS

Instrument must be in “User” mode, see Solution Selection, pg. 11.

Figure 5

°C

KCl

COND

CAL

Advertising