27.6.3. Connecting additional platforms
Connecting 6-wire tensometers
E
- tensometer cable shield
REF+ - “SENSE +” from tensometer (JP1 not soldered)
REF- - “SENSE –“ from tensometer (JP2 not soldered)
IN+
- “OUTPUT+” from tensometer
IN-
- “OUTPUT-“ from tensometer
+5V
- “INPUT+” from tensometer
AGND - “INPUT-“ from tensometer