• Connect the scale to a computer/switch using a PT0017 or PT0014

cable (depending on the Ethernet module on board – see ch. 27.1.2),

• Set the port for communication with a computer – parameter

<P3.1.1 COMP. PORT> set to Ethernet (see ch. 15.1.1),

• Set parameters from the <P2.4 ETHERNET> group (see ch. 14.2):

− Parameter <P2.4.1 WORK MODE> should be set to SERVER,
− Parameters <P2.4.2 IP ADDRESS>, <P2.4.3 SUBNET MASK>,

<P2.4.4 GATEWAY> should be set according to the network
supervisor directions,

<P2.4.5 LOCAL PORT> set the same as in the computer

program (in networks some ports can be taken),

− Parameter <P2.4.8 TIMEOUT> set to 0 in order to prevent

breaking connection.

• Press

the

„Settings” button:

• In the overlap <Transmission setting>:

− Set the communication type „Ethernet Transmission TCP”,
„IP address 1” the same as <P2.4.2 IP ADDRESS> in the scale,
− „Port”/”Com” the same as <P2.4.5 LOCAL PORT> in the scale,

• Confirm the changes by clicking

,

• Rerun the program,

• Press

the

„Connect” button,

• The appropriate communication with scales is signalled by pictogram

in the left part of the display,