28.4.6. Send the result immediately in basic unit

Syntax: SI CR LF

Possible answers:


SI_I CR LF

- command comprehended but cannot be executed at the
moment

SI_A CR LF
MASS FRAME

- command accepted and in progress
- mass value in basic unit is returned


Frame format:

1 2

3

4

5

6 7-15

16 17 18 19 20 21

S I space stability space sign mass space

unit

CR LF


Example:

S I CR LF – computer command
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF - command done, mass value
in basic unit is returned immediately.

28.4.7. Get immediate results from both platforms in basic units

Syntax: SIA CR LF

Possible answers:

SIA_I CR LF

- command comprehended but cannot be executed at the moment

MASS FRAME „P1” CR LF
MASS FRAME „P2” CR LF


- mass values are immediately returned from both
platforms in basic units


Frame format with mass from subsequent platforms as indicator reply:

1 2

3

4

5

6 7-15 16 17 18 19 20 21

P n space stability space

sign

mass

space

unit

CR

LF