Performance specifications – A&D WM:PD4000136 Ambulatory Blood Pressure Monitor User Manual

Page 12

Advertising
background image

Page 10

Functions and Specifications

Specifications

Performance specifications

Measurement method

Oscillometric

Pressurization

Display range

0 ~

320 mmHg

Interval measurement

85 ~

300 mmHg (Fitted)

Manual measurement

185 mmHg (Fixed)

Measurement range

Systolic Blood Pressure

60 ~

280 mmHg

Diastolic Blood Pressure

40 ~

160 mmHg

Pulse rate

30 ~

200 bpm

Accuracy

Pressure

±

3 mmHg

Blood pressure Conforming to 1992 AAMI standard

(

±

3 mmHg or

±

2% measurement whichever is greater)

Pulse rate

±

5 %

Minimum display division Pressure

1 mmHg

Pulse rate

1 bpm

Measurement

Automatic measurement

Manual measurement

Number of measurements Apporx. 180 (Actual measurement number may vary

due to environment and capacity of batteries)

Memory

Up to 300 sets of data

Normal

Clock

Display

During a measurement

Pressure value

After a measurement

SYS, DIA and pulse rate

Error code, function of concealing the measurement data

Clock

24-hours (1997~2096 year, automatic leap year setting)

Batteries

3 x Alkaline battery (type LR6, type AA) or

3 x Ni-Cd battery (type AA)

Type of protection

Internally powered equipment type BF

against electric shock

CE Marking

The label of the medical device

by the EC directive.

0366

C-Tick Marking

The certification trade mark registered

to the ACA by the Trademark office.

AAMI : Association for the Advancement of Medical Instrumentation

ACA: the Australian Communications Authority

CE marking and C-Tick marking are labeled only where they are required.