A&D WM:PD4000136 Ambulatory Blood Pressure Monitor User Manual

Page 35

Advertising
background image

Page 33

Maintenance

Error Code

Error

code

Meaning

Status

Operation and Treatment

E30

E31

E32

E50

E52

E53

E55

E56

E57

E60

E70

E71

E72

E73

E74

E75

E90

E91

Other

Air is exhausted from the

cuff, and an error code is

displayed.

Air is exhausted from the

cuff, and an error code is

displayed.

An error code is displayed.

An error code is displayed

at restarting the product.

An error code is displayed

at restarting the product.

The measurement is

stopped, air is released

from the cuff and an error

code is displayed.

An error code is displayed

at measurement.

Start time is not proper,

interval of last block is not

set in the unit of 120 min.

The error code is displayed

during communications.

This error code is

displayed before the

measurement.

Patient moved during the

measurement.

Monitor code is displayed.

Repair is necessary because of slow

inflation or slow constant exhaust.

Repair is necessary because of slow

constant exhaust.

If you not clear this error, the product

needs repair.

Release the air from the cuff com-

pletely, reset the product. If you not

clear this error, repair is necessary.

The product needs repair.

Replace batteries correctly. If you can

not clear this error, the product needs

repair.

Relax and do not move during the

measurement. If this error occurs

many times, repair is necessary.

Enter parameters for the interval

correctly.

Re-connect the communication cable.

If you can not clear this error, the

product needs repair.

Replace batteries with new ones and

restart communication.

Re-connect the communication cable.

If you can not clear this error, the

product needs repair.

Release the air from the cuff

completely.

Relax and try to quiet during the

measurement. If it occurs in quiet, the

recorder needs repair.

Reset. Turn on power switch again.

Measurement is

greater than 120

seconds.

The constant

exhaust is greater

than 60 seconds.

Clock error.

Pressure offset

error to measure

pulsation.

Memory error.

Battery contact is

defective.

Exhaust error.

Interval setting

error.

RS-232C error.

Low battery for

communication.

Protcol error due

to external

equipment.

Pressure zero

error for safety

circuit.

Safety circuit

detects over load

pressure.

Advertising