A&D WM:PD4000136 Ambulatory Blood Pressure Monitor User Manual

Page 19

Advertising
background image

Page 17

Initializing the recorder

Selection for the Automatic Measurement

Steps for automatic measurement

ex.

First block

0

8:00 ~ 21:59

frequency is 30 minutes

Second block

22:00 ~

0

5:59

frequency is 60 minutes

Third block

0

6:00 ~

0

7:59

frequency is 10 minutes

START STOP

START STOP

START STOP

START STOP

START STOP

START STOP

AUTO ON/OFF

AUTO ON/OFF

AUTO ON/OFF

AUTO ON/OFF

AUTO ON/OFF

AUTO ON/OFF

To next page

Step 1

Press and hold the

START STOP

key for approx. 3 seconds.

The current mode is displayed.

Step 2

Press the

AUTO ON/OFF

key so as to display

for

mode III.

Step 3

Press the

START STOP

key. The mode is stored and the current

start time of the first block is displayed.

Step 4

Press the

AUTO ON/OFF

key so as to display "8" for 8:00 hours

as the start time of the first block.

Step 5

Press the

START STOP

key. The current frequency for the first

block is displayed.

Step 6

Press the

AUTO ON/OFF

key so as to display "30" for 30

minutes as the frequency for the first block.

Step 7

Press the

START STOP

key. The current start time of the

second block is displayed.

Step 8

Press the

AUTO ON/OFF

key so as to display "22" for 22:00

hours as the start time for the second block.

Step 9

Press the

START STOP

key. The current frequency for the

second block is displayed.

Step10

Press the

AUTO ON/OFF

key so as to display " " for 60 minutes

as the frequency of the second block.

Step11

Press the

START STOP

key. The current start time of the third

block is displayed.

Step12

Press the

AUTO ON/OFF

key so as to display "6" for 6:00 hours

as the start time of the third block.

Advertising