Maintenance, Checking accuracy, Start stop – A&D WM:PD4000136 Ambulatory Blood Pressure Monitor User Manual

Page 32: Required equipment

Advertising
background image

Page 30

Maintenance

Checking Accuracy

Required equipment

Accurate office mercury sphygmomanometer or aneroid gauge with inflation

system.

TM-2430 Accuracy Diagnostic Kit (TM2430-90).

A rigid cylinder sized to fit the cuff pressured.

Steps for checking accuracy


Step 1

Turn off the TM-2430 and remove the air hose from the unit.

Step 2

Construct the check system as this drawing.

300
280
260
240
220
200
180
160
140
120
100

80
60
40
20

0

ON

OFF

SYS/INTER

VAL

DIA

PULSE/TIME

AI

R

AUT

O

ON/OFF

STAR

T

STOP

0

1

2

0

1

2

3

4

5

6

7

8

9

A rigid cylinder

Cuff

TM-2430 Accuracy
Diagnostic Kit

Inflator bulb

Exhaust valve and
adjustment screw

Mercury sphyg-
momanometer

Power switch

START STOP switch

Recorder

Step 4

Turn on the power switch, when you press and hold the

START STOP

key.

The TM-2430 blinks the "0" of measurement value.


Step 5

Squeeze the inflator bulb until cuff pressure reaches to 50 mmHg.
Verify that the difference between the blinking display of TM-2430 and

mercury sphygmomanometer is within

±3 mmHg.


Step 6

Squeeze the inflator bulb until cuff pressure reaches to 150 mmHg.
Verify that this difference is within

±3 mmHg.


Step 7

Squeeze the inflator bulb until cuff pressure reaches to 250 mmHg.

Verify that this difference is within

±3 mmHg.


Step 8

Release the cuff air, turn off the TM-2430 and remove the kit.
This blood pressure recorder is a precision instrument. Contact your nearest
A&D office for this inspection, if you need repair.