Display settings

Nio® medical display systems

91

option in the Application Settings dialog is set. See
“Calibration tab” in the description of the Application
Settings dialog.

2. If you have multiple displays connected (multi-head system),

select the display you wish to calibrate by clicking on the
corresponding numbered icon in the Display Settings dialog.

3. Click the Calibrate button.

4. A message appears, showing the different calibration

options.

Figure 45: Calibration choices

5. Select “Using the BARCO LCD sensor” and click OK.

This option is grayed in case NioWatch does not find a
connected measurement device.

6. Follow the guidelines on the screen to complete the

calibration successfully.

Note:

If the Equalization option is set in the Application Settings
and NioWatch finds other displays of the same type that
have been calibrated before, the “Match with reference”
page appears during the calibration process.