3.4 Phase Noise Measurement Tab

In this tab the DUT’s signal frequency and power is detected and the phase noise of the signal in the

specified offset frequency range is analyzed. The measurement principle is based on the phase

discriminator method in combination with a two channel cross-correlation. In general, two internal or

external low noise reference sources are phase locked to the DUT signal. The frequency of the DUT is

detected and a well characterized phase locked loop is closed. Each receiver channel now converts

phase fluctuations between DUT and reference source into low frequency voltage signal by means of

a calibrated phase detector. Inside the PLL bandwidth, the phase differences are suppressed and

must be recovered. The two independent measurement channels are fed into FFT analyzers and

cross-correlator engine to suppress reference and receiver channel noise and to extract DUT phase

noise.

3.4.1

Measurement setup

The following section describes the basic setup procedure to perform the measurement within the

phase noise measurement tab. With just a few configuration steps, the phase noise measurement can

be performed.

Figure 6 Setup section of absolute phase noise tab

Berkeley Nucleonics Corporation 2955 Kerner Blvd., San Rafael, CA 94901

Phone: 415-453-9955, Fax: 415-453-9956, Email: [email protected], Web:

www.berkeleynucleonics.com