2 options, 3 front panel overview – BNC 7000 Series User Manual User Manual

Page 6

Advertising
background image

6

• Universal LAN and USB 2.0 interface

• 24 months calibration cycle

1.2 Options

The following options are available for the SERIES 7000 signal analyzer:

GPIB:

GPIB interface added

1.3 Front Panel Overview

Depending on the SERIES 7000 product configuration, the front panel can contain up to five female

SMA connectors and two female BNC connectors. A green LED indicates the power on/off status of

the instrument. Figure 1 shows the front panels of the SERIES 7300..

Figure 1 Front Panel View of MODEL 7300

DUT/RF IN connector

This female SMA connector is the DUT signal input. The input is AC-coupled

and the impedance is 50 ohm. The damage level is +26 dBm maximum. The maximum allowed DC

level is +/- 10 V.

Channel 1 and 2:

REF IN connector

This female SMA connector is the reference LO input. The impedance is

50 ohm. The damage level is +20 dBm. The maximum allowed DC level is +/- 10 V.

REF OUT connector

(SERIES 7000 ONLY) This female SMA connector is the reference LO

output. The impedance is 50 ohm. In standard operation with internal references the REF

OUT is connected by a short semi-rigid cable to the REF IN.