4 rear panel connections – BNC 7000 Series User Manual User Manual

Page 8

Advertising
background image

8

1.4 Rear Panel Connections

In Figure 2 the rear panels is shown.

Figure 2 Rear Panel View

Channel 1 and 2:

SUPPLY connector (SERIES 7000 and SERIES 7000

only)

This female BNC connector is

the programmable low-noise DC supply voltage output. Output voltage range is 0 V to +5 V,

maximum available current is 80 mA.

BASEBAND IN connector

This female BNC connector is the baseband signal input. To make

baseband noise measurement mode useful for power supply noise measurement, this port is

well protected for DC voltage application. The maximum allowed DC level is +/- 15 V.

LAN connector (8 Pin)

This RJ-45 Host connector is used for connecting via Ethernet with a

controller, such as a PC or Laptop. 10Base-T/100Base-T Ethernet (automatic data rate selection).

USB B connector

Used for connecting with a controller such as a PC or Laptop. Communication is

via USBTMC protocol. Through this port, you can control the SERIES 7000 from external controllers.

For more information on the measurement system using the USB port, see the programmer’s manual.

USB A connector

Used for connecting USB devices such as Memory sticks.

DC IN power r

ece

ptacle

The power receptacle accepts a two-pin plug from the external 6 V DC

power adapter.

GPIB (optional) connector

The connection of an external controller through General Purpose

Interface Bus (GPIB) connector allows you to configure an automatic measurement system.

Berkeley Nucleonics Corporation 2955 Kerner Blvd., San Rafael, CA 94901

Phone: 415-453-9955, Fax: 415-453-9956, Email: [email protected], Web:

www.berkeleynucleonics.com

Advertising