3.4.2

Using External References

Configuring external references requires some additional parameter settings.

A) Select between one channel or two channel measurement.

B) Enter tuning slope (Hz/V) for each measurement channel used. Also enter the tuning

voltage range for each reference. Using the “Meas” button, the tuning slope can also

be measured automatically.

C) If your reference source does not provide sufficient output power, you can also use

internal buffer amplifiers by clicking the “Buffer” checkbox. The buffer amplifier

provides about 10 dB gain with a noise figure of 2 dB.

A)

B)

C)