2.4.2

Connecting to LAN

Connect the instrument to your local area network (LAN) using the Ethernet cable. By default the

instrument is configured to accept its dynamic IP number from the DHCP server of your network. If it is

configured properly, your network router will assign a dynamic IP number to the instrument which will

be automatically displayed on the screen. Your instrument is now ready to receive remote commands.

2.4.3

Direct connectivity to host via Ethernet cable (no router)

You can connect to the instrument to your computer with the Ethernet cable without using a local area

network with DHCP server. To work properly, the network controller (NIC) of your computer must be

set to a fixed IP beginning with 169.254.xxx.xxx (excluding 169.254.1.0 and 169.254.254.255) and

network mask 255.255.0.0 to match the fallback IP that the APPH will assign itself after network

timeout. The APPH will automatically choose an IP address according to ZEROCONF standard.

Therefore, the exact fallback address can’t be predicted.

Connection from a NIC that is configured to use DHCP is also possible. After a preset timeout, the NIC

will assume that no DHCP is available and self-assign a fallback IP that will fall into the range

169.254.xxx.xxx.

Alternatively, you may assign the instrument a fixed IP. Please refer to a later section of this manual to

learn how to do this.

2.4.4

Connecting though USBTMC

Connect the (powered on) instrument to the computer using the USB cable. If properly connected, the

computer host should automatically recognize your instrument as a USBTMC device.

Note if you want to work with the Berkeley Nucleonics GUI, it must be installed with USB support

selected. Then the GUI will detect all attached devices automatically. Open the GUI and follow the

instructions given in Chapter 4.

Alternatively, a VISA runtime environment (NI or Agilent or comparable) must be installed.

Use VISA Write to send the *IDN? Query and use VISA Read to get the response. The USBTMC

protocol supports service request, triggers and other GPIB specific operations.

2.4.5

Installing the BENCHTOP MICROWAVE SOURCE Remote Client

Berkeley Nucleonics’s graphical user interface provides an intuitive control of the instruments. It runs

under Windows™ operating system with minimum requirements. The DLL is embedded in the GUI

application and requires the Microsoft® .NET™ framework to be installed. To install the GUI on the

computer, insert the Berkeley Nucleonics Software and Manual CD into the

CD/DVD

drive. If the setup

doesn’t start automatically double click on the setup.exe to run the auto-installer.

The self-extracting setup provides easy installation and de-installation of the software. The setup

program guides you in a few steps though the installation process. In case the NET framework is not