d) trigger occurrence: single event or repeated

e) trigger type: single point, complete sweep or gated


Confirm settings and arm trigger by pressing the Apply button.

As soon as a sweep is programmed, the output is disabled while the trigger is expected. Upon arrival

of the trigger, the sweep is executed (after given trigger delay). Any trigger signal arriving during

execution of the sweep will be ignored.

If single sweep has been selected, the sweep will be executed only once as soon as first trigger signal

arrives. As soon as the sweep is completed, the frequency will return to the preset CW frequency.

If repeat sweep is selected, the sweep will be executed every time the trigger signal arrives. After the

sweep is completed, the frequency of the RF output is disabled until next trigger arrives.

With trigger type, the user can select the trigger either to execute only a single point from the sweep

or the complete sweep. When gated is selected, the sweep is only executed while the trigger signal is

high.

Figure 3-p TRIGGER Tab

3.10 LF OUT Control

The BENCHTOP MICROWAVE SOURCE rear panel output FUNCT OUT can be programmed using

the LF OUT tab (see

Figure 3-q

). The FUNCT OUT output can be used as