BNC 835 User Manual User Manual

Page 50

Advertising
background image


Berkeley Nucleonics Corporation 2955 Kerner Blvd., San Rafael, CA 94901

Phone: 415-453-9955, Fax: 415-453-9956, Email: [email protected], Web: www.berkeleynucleonics.com

Figure 6-a CW Display with Battery Symbol

Hints for maximizing the battery running time:

1. Fully charge the unit before you use it. Toggle the power switch to ON and then OFF again

while the instrument is powered by the external power adaptor. This will initiate a new charge

cycle.

2. Charging time of a completely discharged battery can be up to 6 hours.

IMPORTANT: The

battery will only be charged when the instrument power switch is in the OFF position.

3. Batteries should always be charged at room temperature. Charging the instrument at very low

or at elevated temperatures may result in early termination of the charging process for

example the battery is not fully charged. For safety reasons the charging does not start when

the internal temperature of the instrument is above 50 °C.

4. Check the battery indicator in the upper right corner of the display. It should indicate full

charge when running on battery power after charging (4 segments).

5. Power consumption of the instrument is reduced when the RF power is switched off thereby

increasing overall battery run time.

6. Battery run time is maximal for ambient temperature between 15 and 25 °C. Self-discharge of

the battery is much faster at temperatures above 30 °C.

7. Avoid storing the instrument in very hot places such as behind the windshield of a car parked

in the sun.

Hints for maximizing battery life expectancy:

1.

The battery will reach its best performance after the first few charge-discharge cycles.

2.

Always use the external power adaptor supplied with the instrument for normal operation and

charging. This will make sure that the charging circuits work as specified.

3.

Fully charge the instrument after running it from the battery for an extended period of time.

4.

If an instrument with internal battery will be stored for a long period of time, fully charge it

before storage, then remove the power adaptor and make sure that the power switch is in the

OFF position. After storage, first charge the unit for 4-6 hours.