6014 & 6016 Operations Manual

Rev C 2/14/12

Page 16 of 56

Connect each incubator to a grounded circuit.
Failure to do so could result in electrical shock.

High concentrations of carbon dioxide can
cause asphyxiation. The use of CO

2

monitors

and alarms is recommended for areas where
CO

2

can collect.

The CO

2

gas supply should be 99.5% pure and

should not contain a siphon tube. Gas pressure
to the unit must be regulated to 15-20PSI.
Failure to do so could cause tubing to burst.


A water inlet fitting

on the back of the unit and ¼” black tubing are provided to connect

the water supply to the incubator. Connect an appropriate water supply to the fitting.
Incoming line pressure should be regulated to not exceed 80 psi.

If a Condensate Recirculator water recycling system was purchased as a water supply,
refer to its user’s manual for proper installation of the water supply.


Connecting a CO

2

supply














The CO

2

supply should be 99.5% and not have siphon tubes. CO

2

pressure should be

regulated to 15-20 psi. CO

2

tank regulators can be purchased through CARON

customer service. Once the cylinder regulator is installed, connect the outlet of the
regulator to the hose barb fitting using the tubing and clamps provided. An inline HEPA
filter is provided to remove any contaminants in the CO2 gas supply. Check the
connections closely for leaks.

If the unit is equipped with a built in gas guard system, there will be 2 gas inlets. Each of
the inlets should be connected to an individual gas tank as described above.

Connecting Electrical Power