Caron 6016 User Manual

Page 35

Advertising
background image

6014 & 6016 Operations Manual

Rev C 2/14/12

Page 35 of 56

the two screws that hold the pen arm be loosened and the pen cartridge and metal pen
arm be removed as an assembly. Unsnap the plastic "U" clip tab of the pen cartridge
from the metal pen arm, remove and discard the old pen cartridge. Replace the new
cartridge by opening the hinge and snapping it securely around the metal pen arm.
Refer to the image below:


Pen Arm Calibration:

To check and/or adjust the recording pen(s) calibration to the outer most temperature
graduation of the chart, press and hold the "change chart" button (#3) until the pen
begins to move off of the chart. Once the pen(s) has moved off of the chart, again press
and hold the "change chart" button (#3) until the pen begins to move back onto the
chart. The pen should briefly stop at the outer most temperature graduation of the chart
before continuing onto the chart to begin recording. If the pen does not stop exactly at
this location on the chart, it can be adjusted or "calibrated" by using the left (#1) or right
(#2) arrow buttons.

When the pen moves back onto the chart and briefly stops, you will have approximately
five seconds in which to adjust the pe

n’s position using the left and right arrow buttons of

Figure 3.

On multiple pen recorders, each pen will move (one at-a-time) onto the chart briefly
stopping at the outer most temperature graduation of the chart at which

time the pen’s

position can be adjusted by using the left (#1) or right (#2) arrow buttons. When the time
to adjust the position of the first pen has expired, the second pen will move onto the
chart briefly stopping at the outer most temperature graduation of the chart at which time
t

he second pen’s position may be adjusted.


Each time the chart paper or fiber tip pen cartridge is changed, you should make sure
that each pen stops at the outer most temperature graduation of the chart paper.
Otherwise, this pen offset will cause the unit to record an incorrect temperature on the
chart.