SECTION 8

– ALARMS


Alarm System Overview

The incubator control system is equipped with an alarm system that constantly monitors
temperature, humidity and CO2 to ensure the user is notified if the cabinet goes into an
alarm condition. Notification occurs via a RED indicator light and an buzzer. Each
alarm condition has been factory programmed to minimize nuisance alarms while
maximizing warning time. The following table contains the alarm conditions being
checked, the factory default alarm range, the amount of time an alarm must be present
to occur (alarm delay), and the message that will be displayed on the individual system
controller.

Alarm Description

Alarm
Deviation

Alarm Delay

Alarm Message

Temp higher than Set-point +1.0

C

15 minutes

Temp Controller

– AL.h1

Temp lower than Set-point

-1.0

C

15 minutes

Temp Controller

– AL.lo

RH higher than Set-point

+10% RH

15 minutes

RH Controller

– AL.h1

RH lower than Set-point

-10% RH

15 minutes

RH Controller

– AL.lo

CO

2

higher than Set-point

+1.0%CO2

15 minutes

CO

2

Controller

– AL.h1

CO

2

lower than Set-point

-1.0%CO2

15 minutes

CO

2

Controller

– AL.lo


In the event an alarm occurs, the alarm indicator will illuminate and an audible alarm will
occur. To temporarily disable the audible alarm, toggle the alarm audible enable switch
to silence. When the alarm condition is corrected both the alarm indicator and the
audible alarm will be disabled. Return the alarm switch to audible.


Changing Alarm Set-points

All alarm set-points were preset at the factory to minimize nuisance alarms that could be
created as a result of door openings. Alarm set-point defaults are shown in the alarm
table earlier in this section. However, alarm set-points can be changed based on
individual user requirements. Each of the controllers are programmed in the same
manner. Press the advance key on the control system that you are changing until either
A.LO1 or A.HI1 is displayed in green. The red displayed value is the deviation from the
set-point that will activate the alarm. Press the UP arrow to increase the deviation,
press the down arrow to decrease the deviation. Press the Infinity Key to exit.