6014 & 6016 Operations Manual

Rev C 2/14/12

Page 43 of 56

If the incubator temperature display reads 40.0

C and the calibrated independent sensor

shows 40.3

C, set the i.CAL offset value to 0.3

C. If the calibrated independent sensor

shows 39.6

C, then the entered offset should be negative. In this example the required

offset to i.CAL would be -0.4

C.