6014 & 6016 Operations Manual

Rev C 2/14/12

Page 45 of 56

Calibrating the CO2

If CO

2

set-point calibration is required, the following steps can be followed:

















Locate the reference instrument’s CO

2

sensor in close proximity to the cabinet’s

geometric center or sample the incubator

’s contents through the sample port located on

the control panel bezel. Be sure that the stabilization times described earlier have been
satisfied prior to performing this calibration. Press the advance key until the green
display reads CAL (calibrate). Pressing the up arrow will increase the CO

2

calibration

offset by 1%. Pressing the down arrow will decrease the CO

2

calibration offset by 1%.

Pressing and holding either button will rapidly scroll the calibration offset. When
finished, press the infinity key to return to the main menu.




Calibrating Optional Chart Recorders

For calibrating the optional front and side mounted chart recorders, refer section 6
(Optional Accessory Operation)

Advance Key

Infinity Key