6014 & 6016 Operations Manual

Rev C 2/14/12

Page 33 of 56

appropriate regulators are not available, contact CARON customer service to purchase
them.

Once the cylinder regulators are installed and adjusted on each tank, connect the outlet
of the regulator on Tank 1 to the hose barb fitting labeled Tank 1 on the back of the unit.
Repeat the process for Tank 2. Turn on the regulated gas supplies and check the
connections closely for leaks.

The power switch is used to turn on/off the operation of the product. If power is off, the
unit will allow gas to flow through Tank 1 if required by the incubator.

The factory default “master tank” is Tank 1. When the appropriate gas pressure is
supplied to both tanks, the master tank will always be used as the gas source. The unit
will swap from the master tank to the alternative tank whenever a low gas pressure
condition is detected.

The alarm switch is used to disable the audible alarm when a low gas condition is
detected. The audible alarm can be silenced while a new tank is installed by selecting
the silence position. Once a low gas pressure condition is resolved, it is recommended
that the alarm switch be returned to audible to ensure audible notification of future low
gas pressure conditions. As a reminder, the Tank 1 and Tank 2 alarm lights will be
illuminated for as long as a low pressure condition is present on the respective tank.



Interior Electrical Outlet (OUTL301)

An optional interior electrical outlet is available to supply power to small interior
appliances such as shakers or stirrers. The outlet is GFI protected. The interior outlet is
fused at 2.0 Amps. It is not intended to power high current draw devices.