Addendum hart communication with the x, Multispectrum ir flame detector – Det-Tronics X3302 Multispectrum IR Flame Detector with HART User Manual

Page 3

Advertising
background image

2.3

95-8613

Digital communication with the X

3302

allows the

operator to monitor the status of the detector, determine

factory settings, adjust field settings, and initiate field

tests. This addendum provides guidance for establishing

HART communication, and describes the HART menu

structure when using the X

3302

with a HART Handheld

Communicator, a PC, or other process interface device

that supports DDL.

NOTE

A minimum level of understanding with regard
to the operation and navigation of the HART
Communicator is required. Refer to the instruction
manual supplied with the HART Communicator for
basic operating instructions.

InTerConneCTIng The hArT CommunICATor
wITh The deTeCTor

Point-to-Point mode
The HART Communicator can connect to the X

3302

at

any wiring termination point in the analog output signal

loop. Connect the HART communicator in parallel with

the X

3302

analog signal or load resistor. The HART

connections are non-polarized.

Important wIrIng note

The HART Communicator does not measure
loop current directly, but instead reads a voltage
signal across a resistance (

250

ohms) in the loop.

The recommended connection point is across
the input impedance of the signal receiver (PLC),
which is a nominal

250

ohms. See Figures

1

to

4

. If testing/programming on a bench, a

250

ohm

load resistor must be used. See Figure

5

.

Switch on the HART Communicator. If a device is found,

the HART Communicator displays the Main menu. If no

device is found, check the connections and verify the

presence of a minimum of

250

ohms load resistance in

series in the loop.

Addendum

HART Communication

with the X

3302

Multispectrum IR Flame Detector

©

detector electronics Corporation 2013

Rev: 10/13

Advertising