SPE-­‐Express  Operation  and  Instruction  Manual  

 

 

www.envexp.com

         

 

 

27  

3.2 Operating A Test

3.2.1

Starting a Test

Press “Start Test”.

Ensure that the filter O-Ring is in place (page 10).

Ensure that the bottle adapter O-Ring is within the groove of the bottle adapter prior to installing the
sample bottle in the unit.
CAUTION: cracking of the glass sample bottle can occur if the bottle O-Ring is not in place.

Once “Start Test” is pressed, a prompt will appear. Press the radio buttons to confirm the following
steps:

o Install a new filter disk mesh side down (The filter cup should be firmly pressed onto the

injection head.)

o Install a sample bottle (The sample bottle should be screwed hand tight into the bottle

adapter.)

o Ensure that the bottle assembly is latched in the upright position (An O-Ring should be

present in the groove of the filter adapter. The filter cup will rest on this O-Ring when the
swinging assembly is in the upright position.)

o Place a new pre-weighed aluminum evaporating pan on the heater block
o Remove the spring clamp then lower the sliding glass chamber into the down position
o Install a new drying cartridge (The drying cartridge should be slid onto the internal nozzle of

the upper glass chamber.)

Then press “Next”.

Once “Next” is pressed, the vacuum will start and a second Pre-Test Checklist will appear as shown,
Test Checklist. Press the radio buttons to confirm the following steps: