1.5.5 Following rules must strictly be adhered to:

Do not operate the centrifuge in case it is not installed correctly.

Do not operate the centrifuge when dismounted (e.g. without housing).

Do not run the centrifuge when mechanical or electrical assembly groups have been
tampered with unauthorized persons.

Do not use accessories such as rotors and buckets, which are not exclusively approved by
HERMLE Labortechnik GmbH, except commercially available centrifuge tubes made of glass
or plastic.

Do not spin extremly corrosive substances, as they may cause material damages and impair
mechanical resistance.

Do not operate the centrifuge with rotors or buckets, which show any signs of corrosion or

mechanical

damage.


The manufacturer is responsible for safety and reliability of the centrifuge, only if:

the unit is operated in accordance with this instruction manual.

modifications, repairs or other adjustments are performed by HERMLE-authorized personnel
and the electrical installation of the related location corresponds to the IEC-regulations.

1.5.6 Warranty

The centrifuge has been subjected to thorough testing and quality controls. In the unlikely case of
any manufacturing faults occurring, the centrifuge and rotors are covered by warranty for a period of

two years from date of delivery. This warranty becomes invalid in case of mishandling, damage and
negligence and further in case of usage of inappropriate spare parts and / or accessories or

unauthorized modification of the unit.



Technical modification rights are reserved by the manufacturer in respect to technical