PRODUCT DESCRIPTION

© Hermle Labortechnik GmbH

Z306 Instruction

manual.docx

7

1.6 Installation of the centrifuge

1.6.1 Unpacking the centrifuge

Model Z 306 is supplied in a carton.
Remove the strap retainer, open the carton and remove the centrifuge. The instruction manual must

always be kept with the centrifuge!

1.6.2 Space requirements

The centrifuge should be installed on an even solid surface, if possible on a laboratory cabinet /
table or some other solid vibration free surface.
During centrifugation, the centrifuge must be placed in a way, that there is a minimum space of
30 cm on each side of the unit according to the standards EN 61010-2-020.
Do not place the centrifuge next to a window or a heater, where it could be disposed to excessive
heat, as the performance of the unit is based on an ambient temperature of 23°C.

1.6.3 Installation

Follow these steps:
• Check whether power supply corresponds with the one named on the manufacturer's rating label,
which is mounted on the rear panel.
• The line voltage circuit braker is max. 10 A (type K) slow release for commonly used instruments.
• In case of emergency, there must be an emergency switch off installed outside the room in order to

disconnect the power supply of the unit.

• Connect the centrifuge with the mains.
(The socket for the power cord must be easy to reach respectively easy to disconnect).
Switching it on using the mains power switch (I).

Open the lid by using the button LID.
• Remove the transport securing device of the motor.