MAINTENANCE

24

Z306 Instruction

manual.docx

© Hermle Labortechnik GmbH

destroy the plastic. The objects must be thoroughly washed up with distilled water after the cleaning
but before the autoclaving. Residues of any cleaning liquids may cause fissures, whitening and
stains.
Gassterilization
Adapters, bottles and rotors may be gassterilized with Ethylenoxyd. According to the duration of the
application you may give long enough an airing to the items after the sterilization and before using
them again.
ATTENTION: Because the temperature may rise during the sterilization, rotors, adapters and
bottles must not be closed respectively must be totally unscrewed

Chemical sterilization
Bottles, adapters and rotors may be treated with the usual liquid disinfectants.
ATTENTION: Before applying any other cleaning resp. Decontamination method than
recommended by the manufacturer, contact the manufacturer to ensure that it will not
damage the unit or the rotor.

3.1.6 Glass breakage

With high g-values, the rate of glass tube breakage increases. Glass splinters have to be removed
immediately from rotor, buckets, adapters and the rotor chamber itself. Fine glass splinters will

scratch and therefore damage the protective surface coating of a rotor. If glass splinters remain in
the rotor chamber, fine metal dust will build up due to air circulation. This very fine, black metal dust
will extremely pollute the rotor chamber, the rotor, the buckets and the samples.
If necessary, replace the adapters, tubes and accessories to avoid further damages. Check the rotor
bores regulary for residues and damages.
ATTENTION: Please check the relevant specifications of the tubes centrifuges with the
manufacturer!

3.2 Service life of rotors, round and rectangular buckets, accessories

Rotors and rotor lid made of aluminium or stainless steel, have a operating life of max. 7 years from
first use.
Transparent rotor lids and caps made of PC or PP as well as rotors, tube racks and adapters of PP
have a maximum service life up to 3 years from first use.
Condition for the operating life:
Proper use, damage-free condition, recommended care.