Table 5: Error messages

Error-No.:

Description

1

Imbalance arose

2

Imbalance sensor is defective

4

Imbalance switch has been activated for longer than 5 seconds

8

Transponder in the rotor is defective

14

CLOSE lid

33

Open lid while motor is running

40

Communication with frequency converter disturbed during start

41

Communication with frequency converter disturbed during stop

42

Short circuit in the frequency converter

43

Undervoltage frequency converter

44

Overvoltage frequency converter

45

Over temperature frequency converter

46

Over temperature motor

47

Over current frequency converter

48

Timeout between control unit and frequency converter

49

Other error frequency converter

55

Overspeed

99

Rotor is not allowed in this centrifuge

FALSE

Inserted rotor does not exist in the program

rotor no

Rotor is not detected

Leap of speed is too big between two measurements