PRODUCT DESCRIPTION

© Hermle Labortechnik GmbH

Z306 Instruction

manual.docx

11

1.7.6 Call up operating data (by skilled personnel or service engineer only!)

In the mode „Basic Adjustments“ you can call up the operating data of the centrifuge. Please
proceed as described under point 1.7.1 to enter this program mode. Press the key „accel/decel“ (5).
In the display „accel/decel“ (A-2) flashes the word „service“.
With the adjusting knob (1) the different information can be called up:
A

= previous starts of the centifuge

H

= previous operating hours

S =

software

version

r

= converter software

E

= list of previous error massage

h

= running time of the motor

The list of the last 99 error messages can be looked over by pressing the key „rpm/rcf“ (4) and scroll
through it by the adjusting knob (1). The respective error codes appear in the display „rpm/rcf“ (A-1).
Please look up in „Table 5: error messages“ (see APPENDIX S. IX).


Here as well you must shortly switch off the centrifuge for changing to the normal program mode

again.


Figure 8