OPERATION

© Hermle Labortechnik GmbH

Z306 Instruction

manual.docx

17

The running time can be pre-selected with the lid open or closed.
To activate the setting of the running time press the key „time“ (6).
In the display „time“ (A-3) flashes the indication „m : s“ or „h : m“, depending on the previous

setting.
To set the desired value use the adjusting knob (1). After exceeding of 59 min 50 sec the indication
changes automatically into „h : m“. After exceeding of 99 hours 59 min the word „cont” appears in
the display „time” (A-3). That continuous run can only be interrupted by pressing the key „stop” (10).
The time countdown as soon as the setted speed is reached.
The display shows always the remaining running time. (see figure 18)
All with number marked passages refer to figure 18

Figure 18

2.3.3 Preselection of brake intensity and acceleration

Through the key „accel/decel“ (5) this function is activated.
By pressing the key once the word „accel“ (M7) flashes in the display „accel/decel“ (A-2). The
desired acceleration can be pre-selected by the adjusting knob (1). The value 0 is equivalent to the

lowest and the value 9 to the highest acceleration.
By pressing the key „accel/decel“ (5) twice, in the display „accel/decel“ (A-2) indicates the word
„decel“ (M8). Now the desired brake intensity can be pre-selected by the adjusting knob (1). The
value 9 is equivalent to the shortest and the value 0 to longest possible brake time.
All with number marked passages refer to figure 19
See „table 4: acceleration and deceleration times“ (APPENDIX P. VIII). There are shown the
acceleration and deceleration times for the acceleration and deceleration stages 0 to 9 for
permissible rotors.