OPERATION

© Hermle Labortechnik GmbH

Z306 Instruction

manual.docx

19

If a program number is already occupied in the display „rpm/rcf“ (A-1) will appear the words „rotor”
(M3) and „22x.xx“ (M4). In case of free program numbers it appears 0.
Close the lid of the centrifuge. Now proceed as already described to set all important run

parameters.
For adaption of data press the key „store“ (12) for approx. 1 second. As a result the word „program“
(M10) disappears. As soon as the key „store“ (12) is no longer anymore, it reappears the word
„programm xx“ (M10) – the (xx) stands for the chosen program place.
If all program numbers are occupied you can take an old number that is not necessary anymore and
just put in the new parameters.
All with number marked passages refer to figure 21.

Figure 21

2.5.2 Recall of stored programs

To recall stored programs press the key „prog“ (11) while the lid is already closed. Inside the display
„time“ (A-3) appears „programm --“(M10). With the adjusting knob (1) you pre-select the desired

program number.
In the respective displays there will appear the stored values for that program.
If there is not the right rotor inside the centrifuge for the pre-selected program, in the display

„rpm/rcf“ (A-1) flashes the word „rotor“ (M3). At the same time the word „FALSE“ and the stored
rotor number „22x. xx“ (M4) will flashing by turns.
All with number marked passages refer to figure 22.

Figure 22